Madgetech Hitemp Datalogger
High Temperature datalogger
Hitemp 140 Datalogger
Madgetech Hitemp Datalogger

Madgetech Hitemp140- Pt(Pointed Tip) Data Logger

Madgetech HiTemp140-PT(Pointed Tip) Data Logger is a submersible, temperature data logger that can work up to 140 °C (284 °F).HiTemp140-PT records date and time stepped readings and has non-unpredictable strong state memory which holds information regardless of whether the battery ends up released

Select Probe Length
Quantity

$979.20
Tax /GST included

Madgetech HiTemp140-PT(Pointed Tip) has a precision of +/ - 0.1 °C (0.18 °F). The HiTemp140-PT comes in two diverse test lengths, both test styles highlight a 24 inch bendable hardened steel test. The HiTemp140-PT-1 has a 1-inch test tip and where the HiTemp140-PT-5 has a 5-inch test tip. The two tests can gauge stretched out temperatures up to 350 °C (662 °F) The test is sturdy and can be spiraled, twisted or calculated toward any path, making it simple to log temperatures inside containers, vials or other difficult to achieve places.

Applications:

• Autoclave Verification, Implement HACCP Programs, Food Preparation and Processing, Environmental Studies, Well Monitoring, Dishwasher Testing, Pasteurization

Highlights:

• NIST traceable
• User Replaceable Battery
• Programmable begin time
• Programmable stop time
• Battery Life Indicator

Advantages:

• Simple Setup and Installation
• Minimal Long-Term Maintenance
• Long-Term Field Deployment

50 Items

Data sheet

Manufacturer
Madgetech
Resolution (Temperature)
0.01 °C (0.02 °F)
Accuracy
Time Accuracy : ±1 minute/month at 20 °C to 30 °C (68 °F to 86 °F) (Stand alone mode)
Dimensions (PT-1 Probe)
Probe tip: 1.7 in x 0.125 in dia. (42 mm x 3.2 mm dia.) || Bendable portion: 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)
Dimensions (PT-5 Probe)
Probe tip: 4.8 in x 0.125 in dia. with 1 in x 0.188 in dia. handle (121 mm x 3.2 mm dia. with 25 mm x 4.8 mm dia. handle) || Bendable portion: 22 in x 0.062 in dia. (559 mm x 1.6 mm dia.)
Interface
IFC400 OR IFC406 USB docking station required; 125,000 baud
Operating Temperature
-40 °C to +140 °C (-40 °F to +284 °F)
Operating Humidity
0 %RH to 100 %RH
Reading Rate
1 reading every second up to 1 reading every 24 hours
Battery Life
1 year typical (1 minute reading rate at 25 °C/77 °F)
Real Time Status
May be used with PC to monitor and record data in real time
Start Modes
• Software programmable immediate start • Delay start up to 18 months in advance
Calibration Date
Automatically recorded within device
Data Format
Date and time stamped °C, °F, K, °R
Stop Modes
Manual or Timed (specific date and time)
Probe Measurement Range
-200 °C to +350 °C (-328 °F to +662 °F)
Temperature Sensor
100 Ω Platinum RTD
Password Protection
An optional password may be programmed into the device to restrict access to configuration options. Data may be read out without the password.
Calibrated Accuracy
±0.1 °C/±0.18 °F (20 °C to +140 °C/68 °F to +284 °F) || ±0.3 °C/±0.54 °F ( -20 °C to +19.99 °C/-4 °F to +67.98 °F) || ±0.4 °C/±0.72 °F (-40 °C to –20.01 °C/-40 °F to -4.02 °F)
Wrap Around
Yes
Memory
32,700 readings
Materials
316 Stainless Steel
Calibration
Digital calibration through software
Software
XP SP3/Vista/Windows 7/Windows 8 (MadgeTech 4 Only)
Approval
CE
Dimensions
(Body) : 1.9 in x 0.97 in dia. (48 mm x 24.6 mm dia.)
Weight
4.2 oz (120 g)