negative air pressure room monitor, room air pressure monitor, clean room pressure differential,pharma dp gauge
water column gauge, room pressure controller, siemens room pressure monitor, pressure measuring devices
negative air pressure room monitor, room air pressure monitor, clean room pressure differential,pharma dp gauge
New

Digital Differential Pressure Indicator Model AI-DP1-1

ACE INSTRUMENTS make Digital Differential Pressure Indicator are digital readout differential pressure indicators designed for measuring the room pressure inside Pharma clean rooms, hospital operation theaters & isolation rooms.

Applications: Pharmaceutical Clean Rooms, Laminar Air Flow Cabinets, HVAC, Bio safety Cabinets, AHUs etc.

Quantity

$145.64
Tax /GST included

Digital Differential Pressure Indicators are part replacement for the analog Magnahelic® gauges & are easy to use with clear 4 digit led display. Differential Pressure Indicators comes with built in sensor & also a buzzer alarm with acknowledgement facility for process violation alert.

Salient features include unit conversion at the press of a key between mm.w.c. & Pascals,integrated buzzer alarm for set point violation, offset provision to meet audit requirements,analog 4-20ma output, 30 mm. Depth for clean room modular wall fitment & RS 485 mod-bus communication for BMS / SCADA / PLC integration.

50 Items

Data sheet

Manufacturer
ACE Instruments
Model
AI-DP1-1-LED
Measuring Range
Select from DropDown
Accuracy
+/-0.5% of Full Scale
Output
4-20 mA Isolated(Optional)
Display
Integrated 4 Digit Red LED 7 Segment Display 0.8”
Maximum Over Pressure
2 Times the rated pressure
Sensors
Integrated High Speed Differential Pressure Sensor
Power 
24 Volts D.C. Typical
Alarms
Built In High & Low Buzzer Alarm for Process Set Parameter Violation with Acknowledgement Key
Construction
M.S. Powder Coated Body with Stainless Steel Front Flush Plate for Modular Wall Fitment in Clean Rooms
Communication
Isolated RS485 Modbus Protocols(Optional)
Connection
2 Nos. Hose Nipples of 4mm. Stainless Steel at the Back
Medium
Air or Gases ( Non Corrosive )
Calibration
Will be provided at no extra cost valid for 1yr traceable to Electronic Test & Development Centre
Dimensions
Instrument Enclosure is in 100×90×30 mm. (W×H×D)
Unit Selection
Factory Fixed Selection for mm.W.C/Pascals