Fluke 725 Multifunction Calibrator

Fluke 725 Multifunction Process Calibrator is a powerful yet easy-to-use field calibrator. Use the measure and source functions to test and calibrate almost any process parameter.

Quantity

$3,820.79
Tax /GST included

Technical Description: 

Measurement Accuracy

Voltage DC

30.000 V

 0.02%+ 2 counts (upper display)

30.000 V

 0.02%+ 2 counts (lower display)

100.00 mV

 0.02%+ 2 counts

-10.00 mV to 75.00 mV

 0.025 % + 1 count (via TC connector)

Current DC

24.000 mA

 0.02%+ 2 counts

Resistance

0.0 to 400.0 Ω

 0.1Ω (4-wire), 0.15Ω (2- and 3-wire)

401 to 1500 Ω

 0.5Ω (4-wire), 1Ω (2- and 3-wire)

1500 to 3200 Ω

 1Ω (4-wire), 1.5Ω (2- and 3-wire)

Frequency

2.0 to 1000.0 CPM

 0.05 % + 1 count

1.0 to 1100.0 Hz

 0.05 % + 1 count

1.00 to 10.00 kHz

 0.05 % + 1 count

Sensitivity

 1 V peak-to-peak-minimum

Pressure

Accuracy

 from 0.025% of range using any of 29 pressure modules.

Modules available for differential, gauge, vacuum, absolute, dual and high pressure.

Source Accuracy

Voltage DC

100.00 mV

 0.02%+2 counts

10.000 V

 0.02%+2 counts

-10.00 mV to 75.00 mV

 0.025 % + 1 count (via TC connector)

Current DC

24.000 mA (Source)

 0.02%+ 2 counts

24.000 mA (Simulate)

 0.02%+ 2 counts

Resistance

15.0 to 400.0 Ω

 0.15 Ω (exc. current 0.15 to 0.5 mA), 0.1 Ω (exc. current 0.5 to 2 mA)

401 to 1500 Ω

 0.5 Ω(excitation current 0.05 to 0.8 mA)

1500 to 3200 Ω

 1 Ω(excitation current 0.05 to 0.4 mA)

Frequency

2.0 to 1000.0 CPM

 0.05%

1.0 to 1100.0 Hz

 0.05%

1.00 to 10.00 kHz

 0.25%

Waveform:

 5 V p-p squarewave, -0.1 V offset



49 Items

Data sheet

Manufacturer
Fluke
Humidity Range
95% (10°C to 35°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
Operating Temperature
10°C to 55°C
Pressure
Accuracy : from 0.025% of range using any of 29 pressure modules
Storage Temperature
-20 °C to 71 °C
Operating Altitude
3000 m
Shock Resistance
30g, 11ms, half-sine shock (or 1meter drop test)
Vibration Resistance
Random, 2g, 5-500 Hz
Approval
EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992
Model
Fluke 725
Dimensions
96 x 200 x 47 mm
Ramp Functions
Source functions: Voltage, current, resistance, frequency, temperature ; Ramps : Slow ramp, Fast ramp, 25% step-ramp
Loop Power Function
Voltage: 24 V ; Accuracy: 10% ; Maximum current : 22 mA, short circuit protected
Step Functions
Source functions: Voltage, current, resistance, frequency, temperature ; Steps ; 25% of range, 100% of range
Warranty
3 years